Product Details

No.258-L IMPACT TESTER (WITH REFRIGERATING MACHINE)

JIS K7110、K7111-1、ASTM D256、D6110、ISO 179-1:2010、180

Specification

Specifications are subject to change without notice.
Temperature Range -35 °C to 60 °C (Refrigerator)
Safety Device Hammer Setting Sensor (Interlock System on the Upper Door), Acrylic Safety Cover
Additional Specifications Refer to the below table
Option Software, Electric Hammer Release Device, Torque Adjustment Device (Only for Izod)
Power Source AC 200 V, 3-Phase, 40 A, 50/60 Hz
Dimensions/ Weight (Approx.) W1,100 × D1,160 × H1,760 mm, 450 kg
Specifications are subject to change without notice.
Test method Charpy Impact Tester Izod Impact Tester
Reference standards JIS K7111, ISO 179, ASTM D6110 JIS K7110, ISO 180, ASTM D256
Capacity 0.5, 1, 2, 4, 5, 7.5, 15, 25 J 1, 2.75, 5, 11, 22 J
Lift-up angle 150° 150°
Impact speed (m/s)* 2.9, 3.8 (3.46) 3.5 (3.46)
Striking Edge Angle* 30 ± 1° (45 ± 2°) 75° (-)
Striking Edge Radius (mm)* R2 ± 0.5 (R3.17 ± 0.12) R0.8 (R0.79 ± 0.12)
Anvil-Edge Distance (mm)* 22 ± 0.2 (22 ± 0.05)
Support Length (mm)* 62 + 0.5 (101.6 ± 0.5)
Accessories Specimen setting gauge, Block gauge, Torque wrench (Izod only) Specimen setting gauge, Block gauge, Torque wrench (Izod only)